Public Lab Research note


Spitfire peak pixel test

by viechdokter | April 26, 2016 22:45 26 Apr 22:45 | #13033 | #13033

What I wanted to do this time is to take a series of spectra of the same CFL lamp in fast succession ("spitfire") at 0.1 seconds intervals over a period of 10 seconds and have a look at the sharpest intensity peaks only.

The questions were: do the peaks move side to side? Do the intensities vary? How much? This is part of my examinations of reproducibility of data using the PLab spectrometer 3.0. I expected to find variations in intensity curves as the lamp was flickering noticably. But will the peaks always stay at the same wavelengths, thus giving a "light fingerprint" of the lamp that is reproducible?

I try to build up on the results of @stoft who found out that single webcam pixels (in my case the peak at a certain wavelength) are not affected by neighbouring pixels at all and that intensity variations in single pixels (wavelengths) occur pretty randomly and not predictably. I also refer to my own earlier noise-by-colour studies.

First the intensities during the first 10 seconds:

For this study I will look at two very distinct peaks - one at 571.1 nanometers and the other at 566-567 nanometers.

571.1 nm curves:

571_nm_peak.jpg

Here are the intensity curves by channel:

Peak_intensity_wobble_at_571_nm.jpg

The blue curve again has the biggest intensity changes. (The green channel was clipped.) In the next diagram I subtracted the minimum values of each channel from the actual values measured each 0.1 second over a period of 10 seconds to get a kind of "base curve":

Peak_intensity_wobble_minus_minimum_values_at_571_nm.jpg

Although red and blue (and of course average) seem to follow roughly the same wobble pattern, the blue curve wobbles much more at this peak.

The 566.1 nm curves:

566_nm_peak.jpg

Peak_intensity_wobble_at_566_nm.jpg

Peak_intensity_wobble_minus_minimum_values_at_566_nm.jpg

Although it is not as clearly visible at this peak the blue curve still has a greater "wobble" than the other channels. But what is more interesting, is that at this peak the overall noise is much higher than at 571.1 nm at the same points in time.

We have different peak intensity changes at different peaks at the same time!

This fits with @stoft s results that each pixel reacts independantly but makes reproducible results for intensities (for example in oil dilution experiment rows) much more difficult to achieve.

Here you can see a comparison between the blue peak intensities at the same points in time at 566.1 and 571.1 nanometers. There is no recognizable pattern between them.

comparison_of_blue_peak_intensities_at_different_peaks.jpg

I don't know what exactly happens that creates such strong differences between between those peaks' intensities but

a) it can't be due to Automatic Gain Control as AGC would not work on single pixels and

b) I think we probably can bury the idea that spectral intensities always relate to ammounts of a certain atom or to a certain type of atom when taking spectra with our PLab 3.0 spectrometer.

The next step will be to assess if the peaks always stay at the same exact wavelengths.

I didn't do this mathematically/statistically but rather "optically" by marking the peak wavelengths per channel in their respective colour on the spreadsheet. Here are screenshots of sections of the different colours with marked maximums:

red_peak_wavelength_a.jpg

green_peak_wavelength_a.jpg

blue_peak_wavelength_a.jpg

These screenshots show that the peaks are pretty stable wavelengthswise. The blue peak is broader than the red peak so you notice a slight wiggle in the red over three or four pixels. Although green is clipped it is always clipped at the exact same place (wavelength).

These are the first hints that concerning the wavelengths the peaks are pretty stable to sideways wiggle.

For anyone wanting to check things, look at the values, curves or data I attach my spreadsheet with all data for the ten seconds spitfire pixel test here:

spitfire_all.ods


1 Comments

Looks like good confirmation of what I observed -- although the clipping does distort the data. Yes, the noise at sample rate is not AGC (that loop bandwidth would be a much lower frequency) so I'd agree is it just noise, mostly Gaussian and largely independent per pixel -- it's just a webcam, not a sensor for an expensive DSLR.

The difficulty with measuring sharp spectral peaks (using 640pix webcams) is that they are operating close to 1pix/nm resolution and the source's true spike bandwidth is much more narrow. This results in the spectrometer's 2-3nm effective bandwidth 'spreading' the narrow source signal some and asking only 1 pixel to define the peak. This is why time-averaging independent pixels helps -- though it cannot add more points. In theory, given a totally stable system and given sufficient number of samples over time, sub-pixel interpolation could be calculated, but that approach wouldn't work across the variety of hardware and configurations. This is also why the PLab 3 paper and Velcro design was so bad -- the mechanical noise of the grating-camera position variation directly contributes to what you term as "sideways wiggle'. Stabilizing those mechanics, at least reduces the error more to just the degree of alignment between the signal peak (optical) and the camera pixel array.

Reply to this comment...